Validation of the cochlear implant artifact correction tool for auditory electrophysiology

Validation of the cochlear implant artifact correction tool for auditory electrophysiology

Authors: 
S. Miller & Y. Zhang
Year: 
2014
Journal: 
Neuroscience Letters
Abstract: 

Auditory event-related potentials (ERPs) collected from cochlear implant (CI) users are often contaminated by large electrical device-related artifacts. Using independent component analysis (ICA), the artifacts can be manually identified and removed, and the ERP responses can be reconstructed from the remaining components. Rao et al., Hear. Res. 268 (2010) 123–132 recently developed an efficient algorithm that uses spatial and temporal statistics of the components to automate CI artifact removal. The purpose of this study was to perform an independent validation of the algorithm. We further assessed whether the ERP responses were stable over the course of one year when analyzed manually or using the semi-automated approach. To achieve these aims, we collected EEG data from 6 adult CI users at two sessions, with one year between each session. We compared their ERP responses reconstructed using the algorithm and the manual approach. We found no significant differences when comparing the two approaches to removing CI artifact across sessions, validating the use of the semi-automated method.